Product Description
Space-saving design with lid opening upward.
Programmable auto-start for initiating a cycle up to 1 week later.
Dual-sensing lid interlock prevents lid opening when chamber is pressurized or when temperature exceeds 97°C.
Automatic air purging by a microprocessor prevents over-pressurization during a cycle.
One-touch lever to easily open and close chamber lid.
Programmable rate of steam exhaust.
Door-closure sensor provides a prompt and the cycle will not start unless the door is closed properly.
In-process display shows status and progress of sterilization cycle.
Agar sterilization mode for sterilizing agar and then automatically maintaining it at liquid temperature.
Unique accessory: Floating sensor to monitor load temperature.
Sterilization timer is activated only when load temperature reaches set temperature.
Agar melting mode for rapidly melting agar and maintaining it at liquid temperature, or for a hot and faster start-up
Memory backup system retains user-programmed parameters even in case of power failure.
Technical Specifications
HV-25 AND HV-50 SPECIFICATIONS
|
MODEL
|
HV-25
|
HV-50
|
| TEMPERATURE | ||
| Operating Range |
105ºC to 126ºC
|
105ºC to 135ºC
|
| Display Range |
128ºC
|
137ºC
|
| Warming Range |
45ºC to 60ºC
|
|
| Agar Dissolution |
Heating at 60ºC to 100ºC for up to 1 hour (temperature and time programmable), followed by 45ºC to 60ºC
(programmable for up to 20 hours to prevent coagulation) |
|
| GAUGE PRESSURE | ||
| Operating Range |
3 to 22 psi gauge (0.02 to 148MPa)
|
3 to 30 psi (0.02 to 0.206MPa)
|
| Maximum Allowable |
27 psi (0.186MPa)
|
38 psi (0.26MPa)
|
| TIMER | ||
| Sterilization |
1 to 250 minutes
|
|
| Delayed Auto-start |
Program autoclave to start 1 minute to 7 days later
|
|
| DIMENSIONS | ||
| Overall (W x D x H) |
19.0″ x 18.0″ x 37.4″
(48 cm x 46 cm x 95 cm) |
20″ x 21″ x 41″
(51 cm x 53 cm x 104 cm) |
| Chamber (Diameter x Height) |
25 liters
9.4″ x 21.6″ (24 cm x 55 cm) |
50 liters
11.8″ x 28.0″ (30 cm x 71 cm) |
| SAFETY DEVICES |
Dual-sensing lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, microprocessor-controlled pressure relief valve, lid-closure detector, over-current detector and circuit breaker
|
|
| PROGRAMS | ||
| 1. For agar |
Heating -> Sterilization -> Exhaust (from no exhaust to slow, programmable) -> Warming -> Completion
|
|
| 2. For liquids |
Heating -> Sterilization -> Exhaust (from no exhaust to slow, programmable) -> Completion
|
|
| 3. For solids |
Heating -> Sterilization -> Exhaust (fast, default) -> Completion
|
|
| 4. For melting agar or for faster start-up |
Rapid melting at 60°C to 100°C (temperature and time programmable) -> Maintain at 45°C to 60°C (temperature programmable) -> Completion
|
|
| ACCESSORIES | ||
| Stainless steel |
Mesh wire baskets, buckets (pails), and round test tube racks
|
|
| Factory or field installed |
Floating (load) sensor, digital printer, and data logger
|
|
| POWER | ||
| Source |
120V, 50/60 Hz (220V/240V available)
|
|
| Consumption |
1500 watts (1500VA)
|
2000 watts (2000VA)
|
| SHIPPING WEIGHT |
160 lb (73 kg)
|
180 lb (82 kg)
|
WARRANTY: Autoclaves are protected by a 12-month warranty.
HV-85 AND HV-110 SPECIFICATIONS
|
MODEL
|
HV-85
|
HV-110
|
| TEMPERATURE | ||
| Operating Range |
105ºC to 135ºC
|
|
| Display Range |
137ºC
|
|
| Warming Range |
45ºC to 60ºC
|
|
| Agar Dissolution |
Heating at 60ºC to 100ºC for up to 1 hour (temperature and time programmable), followed by 45ºC to 60ºC
(programmable for up to 20 hours to prevent coagulations) |
|
| GAUGE PRESSURE | ||
| Operating Range |
3 to 30 psi (0.021 to 0.206MPa)
|
|
| Maximum Allowable |
38 psi ( 0.26MPa)
|
|
| TIMER | ||
| Sterilization |
1 to 250 minutes
|
|
| Delayed Auto-start |
Program autoclave to start 1 minute to 7 days later
|
|
| DIMENSIONS | ||
| Overall (W x D x H) |
26.0″ x 25.6″ x 39.4″
(66 cm x 65 cm x 100 cm) |
26.0″ x 25.6″ x 46.5″
(66 cm x 65 cm x 118 cm) |
| Chamber (Diameter x Height) |
85 liters
16.5″ x 24.2″ (42.0 cm x 61.5 cm) |
110 liters
16.5″ x 31.3″ (42.0 cm x 79.5 cm) |
| SAFETY DEVICES |
Dual-sensing lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, microprocessor-controlled pressure relief valve, lid-closure detector, over-current detector and circuit breaker
|
|
| PROGRAMS | ||
| 1. For agar |
Heating -> Sterilization -> Exhaust (from no exhaust to slow, programmable) -> Warming -> Completion
|
|
| 2. For liquids |
Heating -> Sterilization -> Exhaust (from no exhaust to slow, programmable) -> Completion
|
|
| 3. For solids |
Heating -> Sterilization -> Exhaust (fast, default) -> Completion
|
|
| 4. For melting agar or for faster start-up |
Rapid melting at 60°C to 100°C (temperature and time programmable) -> Maintain at 45°C to 60°C (temperature programmable) -> Completion
|
|
| ACCESSORIES | ||
| Stainless steel |
Mesh wire baskets, buckets (pails), and round test tube rack
|
|
| Factory or field installed |
Floating (load) sensor, digital printer, and data logger
|
|
| POWER | ||
| Source |
208V/220V/240V, 50/60 Hz, single phase
|
|
| Consumption |
3000 watts (3000VA)
|
4000 watts (4000VA)
|
| SHIPPING WEIGHT |
282 lb (128 kg)
|
308 lb (140 kg)
|
WARRANTY: Autoclaves are protected by a 12-month warranty.













Reviews
There are no reviews yet.