Technical Specifications
Est. Shipping Width Inches (mm) | 31 (787) |
Est. Shipping Height Inches (mm) | 49 (1245) |
Est. Shipping Depth Inches (mm) | 33 (838) |
Access Port Location | Right Side |
Air Cleanliness (ISO 14644) | ISO-Class-5 |
Chamber Air Circulation Method | Air Pump |
Chamber Air Filter | 99.97% Efficiency @ ≥ 0.3 Micron Capsule Filter Located in Sensor Bay |
Chamber Heater(s) Location | Bottom of Chamber |
Chamber Jacket Technology | Water Jacket, 20 Gallon (75.7L) |
CO2 Accuracy | ± 0.1% |
CO2 Control Logic | Fixed Algorithm/Manual, Environmental Adaptable |
CO2 Display Resolution | 0.1% |
CO2 Range | 0.1 to 20%, 5% Default Set Point |
CO2 Recovery | Up to 5% ± 0.2% / -0.5 in 5 minutes Average |
CO2 Sensor Type | Infrared (NDIR) |
Control System | NuTouch |
Data Output Features | RS-485, USB, 4-20mA |
Door and Perimeter Heater Control Logic | Base Duty Cycle set to chamber temperature Set-Point, Proportional 0-100% [manually adjustable |
Electrical Configuration | 115V50Hz, 230V5060Hz |
Exterior Depth Inches (mm) | 27 (685) |
Exterior Door Seal | Magnetic Outer Door Gasket |
Exterior Height Inches (mm) | 37.75 (958) |
Exterior Material | 16 Gauge Steel with Textured Polyurethane Finish |
Exterior Width Inches (mm) | 25.625 (649) |
Footprint Depth inches (mm) | 17.5 (445) |
Footprint Width inches (mm) | 22 (557) |
In-line Gas Filters | 99.97% Efficiency @ ≥ 0.3 Micron Disk Filters for Gas & Air Supplies, Disk Filter for CO2 Sensor |
Segmented Doors | Optional Set of 4 Inner Doors |
Inner Doors | One Standard, Sectional optional |
Insulation | Insulation with R5.0 Rating |
Interior Depth Inches (mm) | 20.625 (524) |
Interior Door Seal | Silicone Rubber Gasket |
Interior Height Inches (mm) | 24 (611) |
Interior Material | Polished Type 304L 16 Gauge Stainless Steel |
Interior Volume Cubic Feet (Liters) | 5.65 (160) |
Interior Width Inches (mm) | 20.375 (518) |
Net Weight lbs (kg) | 403 (183) |
O2 Accuracy | ± 0.25% |
O2 Control Logic | Fixed algorithm/manual environmental adaptable |
O2 Display Resolution | 0.1% |
O2 Measuring Range | 0 - 25% |
O2 Set Point Range | 0.5 -21.0% (ambient) (21.0% Default) |
O2 Recovery | 5.0 ± 0.5% twenty minutes on Average |
O2 Sensor Type | Zirconia Ceramic |
Power Cord Length Foot (Meter) | 8 (2.5) |
Remote Alarm Terminals | Standard |
RH Accuracy | N/A |
RH Control Method | Water Pan |
RH Control Logic | N/A |
RH Control Range | N/A |
RH Display Resolution | N/A |
RH Evaporator Box Capacity | N/A |
RH Recovery | N/A |
RH Sensor Type | N/A |
RH Water Pan Capacity | 2L (Recommended Fill 1.5L) |
Sample Port Location | Inner Glass Door |
Shelf Capacity lbs (kg) | 25 lbs per Shelf / 125 lbs per Incubator |
Shelf Depth Inches (mm) | 18 3/4 (476) |
Shelf Material | Stainless Steel |
Shelf Width Inches (mm) | 18 (457) |
Shelves | 4 Shelves Supplied, 16 Shelves Maximum Capacity |
Est. Shipping Weight lbs (kg) | Estimated 370 lbs (168 kgs) |
Decontamination Cycle | Not available |
Temperature Accuracy °C | ± 0.1% |
Temperature Control | Water Jacket |
Temperature Control Range | 5°C above ambient (to a 30°C max ambient) to 55°C |
Temperature Display Resolution | 0.1°C |
Temperature Recovery | 0.12°C/min. on Average |
Temperature Sensor Type | Precision Integrated Circuit |
Temperature Set-Point Range | 5°C to 55°C (37.0°C Default) |
Temperature Uniformity °C | ± 0.20°C @ 37°C |
Water Pan Depth | 1.5 (38) |
Water Pan Length Inches (mm) | 12 (305) |
Water Pan Width | 10 (254) |