Technical Specifications
Est. Shipping Width Inches (mm) | 31 (787) |
Est. Shipping Height Inches (mm) | 51(1219) |
Est. Shipping Depth Inches (mm) | 31 (787) |
Access Port Location | Right Side |
Air Cleanliness (ISO 14644) | ISO-Class-5 |
Chamber Air Circulation Method | Air Pump |
Chamber Air Filter | 99.97% Efficiency @ ≥ 0.3 Micron Capsule Filter Located in Sensor Bay |
Chamber Heater(s) Location | Walls, Top and Bottom of Chamber |
Chamber Jacket Technology | Air Jacket via One or Two Axial Fans Drawing Air into Shell |
CO2 Accuracy | ± 0.1% |
CO2 Control Logic | Fixed Algorithm/Manual, Environmental Adaptable |
CO2 Display Resolution | 0.1% |
CO2 Range | 0.1 to 20%, 5% Default Set Point |
CO2 Recovery | Up to 5% ± 0.2% / -0.5 in 5 minutes Average |
CO2 Sensor Type | Infrared (NDIR) |
Control System | NuTouch |
Data Output Features | RS-485, USB, 4-20mA |
Door and Perimeter Heater Control Logic | Base Duty Cycle set to chamber temperature Set-Point, Proportional 0-100% [manually adjustable |
Electrical Configuration | 115V50Hz, 230V5060Hz |
Exterior Depth Inches (mm) | 27.188 (691) |
Exterior Door Seal | Magnetic Outer Door Gasket |
Exterior Height Inches (mm) | 39.688 (1008) |
Exterior Material | Cold Rolled 16 & 18 Gauge Steel, Finished in Powder Coated Polyurethane Finish |
Exterior Width Inches (mm) | 26.750 (680) |
Footprint Depth inches (mm) | 20.75 (527) |
Footprint Width inches (mm) | 23.75 (603) |
In-line Gas Filters | 99.97% Efficiency @ ≥ 0.3 Micron Disk Filters for Gas & Air Supplies, Disk Filter for CO2 Sensor |
Segmented Doors | Optional Set of 6 Inner Doors, 2 Across by 3 Down |
Inner Doors | One |
Insulation | High Density Insulation with High "R" Rating |
Interior Depth Inches (mm) | 20.053 (509) |
Interior Door Seal | Silicone Rubber Gasket |
Interior Height Inches (mm) | 28.500 (724) |
Interior Material | Polished Type 304L 16 Gauge Stainless Steel |
Interior Volume Cubic Feet (Liters) | 7.06 (200) |
Interior Width Inches (mm) | 21.257 (540) |
Net Weight lbs (kg) | 235 (106) |
O2 Accuracy | ± 0.25% |
O2 Control Logic | Fixed algorithm/ manual environmental adaptable |
O2 Display Resolution | 0.1% |
O2 Measuring Range | 0 - 25% |
O2 Set Point Range | 0.5 - 21.0% (ambient) (21.0% Default) (2 - 21.0% when RH setpoint is above 90%) |
O2 Recovery | 5.0 ± 0.5% twenty minutes on Average |
O2 Sensor Type | Zirconia Ceramic |
Power Cord Length Foot (Meter) | 8 (2.5) |
Remote Alarm Terminals | Standard |
RH Accuracy | +5% /- 3% |
RH Control Method | Water Pan or Evaporator Box, Depending on Model |
RH Control Logic | Fixed algorithm/ manual environmental adaptable |
RH Control Range | Ambient dependent to 95% |
RH Display Resolution | 0.1% |
RH Evaporator Box Capacity | 3L |
RH Recovery | 90% +5%/-3% within 25 minutes on average |
RH Sensor Type | Precision capacitance bulk-polymer |
RH Water Pan Capacity | 2L (Recommended Fill 1.5L) |
Sample Port Location | Inner Glass Door |
Shelf Capacity lbs (kg) | 25 lbs per Shelf / 125 lbs per Incubator |
Shelf Depth Inches (mm) | 18 3/4 (476) |
Shelf Material | Stainless Steel |
Shelf Width Inches (mm) | 19 3/4 (502) |
Shelves | 4 Shelves Supplied, 20 Shelves Maximum Capacity |
Est. Shipping Weight lbs (kg) | Estimated 325 lbs (147 kgs) |
Decontamination Cycle | True (False on NU-5800, NU-5800E) |
Temperature Accuracy °C | ± 0.1% |
Temperature Control | Direct Heat |
Temperature Control Range | 5°C above ambient (to a 30°C max ambient) to 55°C |
Temperature Display Resolution | 0.1°C |
Temperature Recovery | 0.12°C/min. on Average |
Temperature Sensor Type | Precision Integrated Circuit |
Temperature Set-Point Range | 5°C to 55°C (37.0°C Default) |
Temperature Uniformity °C | ± 0.35°C @ 37°C |
Water Pan Depth | 1.5 (38) |
Water Pan Length Inches (mm) | 12 (305) |
Water Pan Width | 10 (254) |