Product Description
- GMI Certified – warranties up to 1 year
- Proof of Performance below (& included with purchase)
- Computer & Software Included
Shimadzu’s Cannabis and Hemp Testing Solutions cover a broad range of applications, from accurate potency testing and terpene profiling to contaminate testing for pesticides, residual solvents, heavy metals, and mycotoxins/aflatoxins. GMI has re-certified this instrument to factory standards and can supply the instruments, methods and experience necessary to get you up and running quickly. Shimadzu scientific instruments have been applied to natural products, food sciences, pharmaceuticals and environmental testing for decades. Whether you are a chromatography expert or a novice, our platforms and training ensure that your laboratory will generate quality results from a wide variety of cannabis samples. Our easy to use platforms save you time and money, maximizing the profits of your laboratory. GMI can help you with your QA/QC testing, peak harvesting analysis and cannabis research.
Proof of Performance
Unit | Inspection Item | Method | Criteria | Result | Judgement | ||
Controller | ROM, RAM Check | Self-Diagnosis | ROM, RAM OK | – | Pass | ||
Degassing Unit | Output Voltage | Measuring Vacuum Output Voltage | 100+2mV | 99 mV | Pass | ||
Pump
|
Flow Rate Accuracy | at 1.0mL/Min | 1+0.010mL/Min | Pass | |||
Flow Rate Stability | at 1.0mL/Min | <0.05 MPa | 0.02 MPa | Pass | |||
Pressure Limiter | P.MIN, P. MAX | Pass is displayed | – | Pass | |||
Gradient Concentration Accuracy
|
Flow Line | 10% | 50% | 90% | Pass | ||
Flow Line A-B | -0.14 | -0.15 | -0.13 | Pass | |||
Flow Line A-C | 0.15 | 0.26 | 0.24 | Pass | |||
Flow Line A-D | -0.17 | -0.10 | -0.16 | Pass | |||
In All Cases +0.5% | Pass | ||||||
Auto Sampler
|
Injection Volume Repeatability
|
5 uL(n=6) | %RSD <0.2% |
0.06% |
Pass
|
||
1 uL(n=6) | %RSD <0.5% |
0.05% | |||||
Injection Volume Accuracy | 50uL Injection | 1.0% + | -0.01% | Pass | |||
Carry-Over | Sample: Caffeine | <0.0025% | 0.00% | Pass | |||
Sample Cooler Temp Accuracy | Temp. at the bottom of 1.5mL vial (4℃) | 4.0+3.0℃ | R: 5.1℃ | Pass | |||
Column Oven
|
Temperature Accuracy | 40℃ | 40+0.8℃ | 40.0℃ | Pass | ||
Temperature Fluctuation | 40℃ | Fluctuation <10℃ | 0.01℃ | Pass | |||
UV-Vis Photodiode Array Detector
|
Absorbance Linearity | Caffeine Solution | 5.0%+ | -1.40% | Pass | ||
Wavelength Accuracy
|
241.5nm | 1.0nm+ | -0.1nm | Pass | |||
656.1nm | 1.0nm+ | -0.1nm | Pass | ||||
Lamp Energy | D2 Lamp (210~240nm) | >7.0V | 8.0V | Pass | |||
Drift (ASTM Method) | 250nm | <5.0×10^-4 AU/hr | 1.5 | Pass | |||
Noise (ASTM Method) | 250nm | <6.0×10^-6 AU | 3.3 | Pass | |||
Absorbance Accuracy | 245nm Reference Value | 3.0%+ | -1.10% | Pass | |||
Total Judgement | Pass |
Reviews