Product Description
The Beckman Optima™ TLX ultracentrifuge, is a tabletop model designed for a variety of high-speed applications. Along with operator safety, the Beckman ultracentrifuge system prioritizes maximum performance capabilities with a force of 657,000 x g and sample volumes ranging from 0.2 mL to 5.1 mL per tube.
The Beckman Optima™ TLX centrifuge offers enhanced operator convenience. Equipped with an easy to use control panel, this Beckman centrifuge also has a temperature control system, a self-purging vacuum system and a programmable memory that can store up to 10 different profiles and more.
The control panel of this Optima™ ultracentrifuge system is located on top of the machine for convenience and easy access and visibility. During a run, the actual condition is shown in the control panel including speed, temperature, and time. The thermoelectric temperature control system uses forced air and doesn’t require any coolant.
The Beckman Optima™ TLX centrifuge also comes with safety features. The centrifuge has imbalance-tolerant, direct vacuum-encased, air-cooled drives, as well as a steel chamber door which has an electromechanical door locking mechanism. The door automatically locks when manually closed.
Beckman Optima TLX Ultracentrifuge System Specifications
Height | 53.3 cm (21 in) |
Depth | 58.4 cm (23 in) |
Weight | 81 kg (178 lb) |
Approximate Acceleration Time | 10 acceleration profiles–9 slow rates from 0 to 5,000 rpm followed by full acceleration to set speed; or maximum acceleration |
Approximate Deceleration Time | 11 deceleration profiles–10 slow rates from 5,000 to 0 rpm, including coasting to a stop without brake; or full dynamic braking from set speed |
Electrical Requirements | 50 Hz, 198 to 254 VAC, 6 A |
Maximum Heat Dissipation into Room under Steady-State Conditions |
0.5 kW (1,750 Btu/hr) |
Noise Level 1 Meter in Front of Centrifuge |
< 60 dBA |
Set Speed | Actual rotor speed ± 50 rpm of set speed |
Set Temperature | 2 to 40°C in 0.1° increments |
Set Time | to 99 hr 59 min |
Speed Range | 5,000 to 120,000 rpm |
Temperature Control | ± 2°C of set temperature (after equilibration) |
Ambient Temperature Range | 15 to 35°C ambient |
Clearances Required | 7.6 cm (3.0 in) both sides and rear |
Humidity Restrictions | < 95% (noncondensing |
Pollution Degree | 2 |
Time – Actual Display | Indicates run time remaining |
Call us today to learn more about the Beckman Optima™ TLX Ultracentrifuge
user manual
Reviews