Product Description
- #1 Scale up Pilot FPLC in the world
- Hygienic design for versatile large scale purification
- Huge dynamic capacity
- Assisted validation with UNICORN software
- GMI engineers are factory trained by the OEMs & avg. 22 years of experience
- Simple cleaning validation: Change wet parts easily
- GMI Certified Pre-Owned
- Eligible for a 1-Year Extended Warranty
The GE Amersham Pharmacia AKTA pilot is a dedicated automated liquid chromatography (LC) system designed to carry out development, scale-up, and production processes. This bio-compatible, hygienic and sterilizable system can purify products in milligrams of up to tens of grams. As the gold standard in the industry, the GE Amersham Pharmacia AKTA pilot LC system meets all good laboratory practice (GLP) and current good manufacturing practice (cGMP) guidelines which are required in Phase I-III of drug development and final-scale production.
The GE Amersham Pharmacia AKTA pilot is a complete system which contains all fluid handling components necessary to perform smooth delivery and mixing, sample application as well as in-line monitoring.
Pre-programmed method templates and optimized system configurations minimize preparation time by allowing rapid media screening, method scouting and optimization as well as scale up experiments. Dedicated method templates are supplied for the most common scouting procedures including automatic media screening and automatic eluent pH scouting for ion exchange chromatography. Convenient sample application and column packing is also possible by regulating the flow rate, while pressure can also be monitored through four points. The UNICORN software displays the monitoring of conductivity, UV and pH. Additional screens can be connected to the system for special purposes. The GE Amersham Pharmacia AKTA pilot is equipped with sanitary sensors allowing continuous in-line measurement of pH, conductivity and UV absorbance. These detectors deliver accurate and reliable monitoring through self-test and self-calibration. The flow cells are placed close together and are easily accessible from the front panel to facilitate maintenance. As a result, band broadening and time delay between detectors are lessened. Also known as: pilot, Amersham, akta, pilot, fplc, aktapilot, akta-pilot, pharmacia pilot.
Let our team of experts help you find the perfect equipment for your application and budget, schedule a free meeting now:
Specifications
| Isocratic and gradient flow rate | 4–400 mL/min |
| Double mode flow rate | Up to 800 mL/min |
| Operating pressure | Up to 2 MPa (20 bar, 290 psi) between system pump and column inlet |
| UV wavelength range | 190–700 nm in steps of 1 nm, 206–700 nm with AKTApilot cell, 3 wavelengths simultaneously |
| pH range | 0–14 (spec. valid between 2–12) |
| Conductivity range | 1 µS/cm to 999.9 mS/cm |
| Holdup volume | Up to 31 ml |
| Max backpressure | less than 0.06 MPa before column/ 400 ml/min less than 0.045 MPa after column /400 ml/min |
| Solvent compatibility | All commonly used chromatographic solvents |
| Degree of protection | IP 24 |
| Dimensions | 29.5″ W x 21.3″ D x 35″ H (75 × 54 × 90 cm) |
| Weight | 251 lbs. (114 kg) without arm |
| Power | 100-240 V 50-60Hz; 800 VA power max |