Replacement Parts & Accessories
To view available parts for this instrument and instruments like it, click here.
Our parts and accessories inventory is constantly changing. Don't see what you're looking for? Give us a call, we'll be happy to help.
The future of flame photometry has arrived!
The “Flash” BWB Flame Photometer mini product line was created in January 2018 to answer the request from global distributors and users for a lower cost version of the Highly Acclaimed BWB “XP” that has dominated the Flame Photometry market since 2007.
These new models feature the identical inbuilt custom designed air compressor (rated for thousands of hours of trouble free service), the expandable Menu structure for units of measure, and the customer benefit of the “Just Add Gas” philosophy for simplified ordering and start up. These are the same benefits as our premium line of XP Flame Photometers, just with a slightly restricted Options list and fewer Laboratory and Fluids kit items.
On board “Level Sensor” for drain cup alert
Intuitive user interface for true ease of use – Display prompts step by step operation
Built-in Air Compressor
Solutions and labware included
6 User selectable units of measure: PPM, MG/L, MEQ/L, %, MG/KG, MMOL/L
1-Year factory warranty included – Extended 5-Year Warranty Available
Sample rate
2 – 3.5ml/min
Tubing materials
Silicone and Tygon®
Required desk space for Instrument
50cm (H) x 45cm (W) x 45cm (D)*
For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of heat from the chimney.
Instrument size
XP Flame Photometer
51cm (H) x 38cm (W) x 41cm (D)
(20in x 14in x 16in)
Shipping
62cm (H) x 55cm (W) x 47cm (D)
(24in x 22in x 19in)
Weight
Flash Flame Photometer – 15.3kg (33.7lbs)
Shipping – 22kg (48.5 lbs)
Optimal range:
Single point calibration
Na 0.1 – 60ppm
K 0.05 – 100ppm
Li 0.05 – 30ppm
Ca 1.0 – 100ppm
Multi Point Calibration
Na 0.1 – 1000ppm
K 0.05 – 1000ppm
Li 0.05 – 1000ppm
Ca 1.0 – 1000ppm
Reproducibility
<1% Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100ppm or less. (After instrument stabilisation).
Specificity
Na/K/Li = <0.5% to each other when equal in concentration at <100ppm
Limit of detection (LOD) and
limit of quantification (LOQ)
LOD LOQ
Na – 0.03ppm Na – 0.1ppm
K – 0.02ppm K – 0.05ppm
Li – 0.02ppm Li – 0.05ppm
Ca – 0.3ppm Ca – 1ppm
Time to stability
Less than 15 seconds after sample is introduced into the flame.
Ions measured
Simultaneous measurement of Na, K, Li, Ca
Interfaces
USB
0-2.5 volt output (based on sample concentration linked to element of users choice)
Optional 4-20mA output in place of the above
Optional integrated printer
.csv and .pdf generated reports and files via FP-PC software
Recommended minimum warm up time based on ambient temperature
21ºC – 40 minutes
Power requirements
100V – 250V AC at 50 or 60Hz automatically selecting
Fuel requirement
Propane, Butane or Natural Gas* regulated to 19Bar. Flow rate of 0.4l/min
*with modifications. BWB Technologies recommends either Propane or Butane for optimum results.
Readout
LCD, four line, alpha numeric, back lit.
The design criteria for this new 2018 line was to provide a modernized replacement flame photometer for the hundreds of units in service in the USA and Canada that are now decades old. These older units have no replacement parts and no servicing is currently available. These are Perkin Elmer, Jenway Instruments, Corning and Sherwood Instruments units as well as the units supplied by Cole Parmer, VWR and Fisher Scientific.
When the costs of adding the extra components that are required to operate the flame photometers sold by our competition are totalled up, (calibration solutions, diluent solutions, various user replacement optional detection filters, optional software that ties to computers; AND the big cost, the required Air Compressor) the benefit of simplicity and cost savings of the “Just Add Gas” – “only one order number needed” we provide is apparent.
Simple to order, simple to use, minimal table top space required, Major Power in simple software as standard (FP-PC), the new “FLASH” is another exciting offering from BWB Technologies USA.
For new emerging markets: aqua farming, flow injection analyses, educational markets as well as the Replacement Flame Photometer marketplace in the USA.
To view available parts for this instrument and instruments like it, click here.
Our parts and accessories inventory is constantly changing. Don't see what you're looking for? Give us a call, we'll be happy to help.
Service Agreements |
Silver |
Gold |
First Class Ship & Fix |
Economy Ship & Fix |
Preventative Maintenance Visit | 1 Preventative Maintenance Visit per Year | 1 Preventative Maintenance Visit per Year | ||
Parts Covered | ||||
Labor Covered | ||||
Unlimited Remote Tech Support | ||||
Depot Repair | ||||
Depot Preventative Maintenance |