Replacement Parts & Accessories
To view available parts for this instrument and instruments like it, click here.
Our parts and accessories inventory is constantly changing. Don't see what you're looking for? Give us a call, we'll be happy to help.
Designation | UFV700-230V | UFV700-230V-W | UFV700UL-240V | UFV700UL-120V |
Temperature range | -90…-40 °C | -90…-40 °C | -90…-40 °C | -90…-40 °C |
Temperature uniformity at -80°C | 2.5 ± K | 2.5 ± K | 2.5 ± K | 2.5 ± K |
Temperature fluctuation at -80°C | 1.5 ± K | 1.5 ± K | 1.5 ± K | 1.5 ± K |
Designation | UFV700-230V | UFV700-230V-W | UFV700UL-240V | UFV700UL-120V |
Rated Voltage | 230 V | 230 V | 208…230 V | 115 V |
Power frequency | 50 Hz | 50 Hz | 60 Hz | 60 Hz |
Nominal power | 1,6 kW | 1,6 kW | 1,8 kW | 1,4 kW |
Unit fuse | 10,0 A | 10,0 A | 10,0 A | 15,0 A |
Phase (Nominal voltage) | 1~ | 1~ | 2~ | 1~ |
Designation | UFV700-230V | UFV700-230V-W | UFV700UL-240V | UFV700UL-120V |
Interior volume | 700 L | 700 L | 700 L | 700 L |
Net weight of the unit (empty) | 288 kg | 288 kg | 288 kg | 288 kg |
Permitted load | 200 kg | 200 kg | 200 kg | 200 kg |
Permitted load of individual compartment | 50 kg | 50 kg | 50 kg | 50 kg |
Wall clearance back | 100 mm | 100 mm | 100 mm | 100 mm |
Wall clearance sidewise | 250 mm | 250 mm | 250 mm | 250 mm |
Designation | UFV700-230V | UFV700-230V-W | UFV700UL-240V | UFV700UL-120V |
Width net | 1110 mm | 1110 mm | 1110 mm | 1110 mm |
Height net | 1966 mm | 1966 mm | 1966 mm | 1966 mm |
Depth net | 938 mm | 938 mm | 938 mm | 938 mm |
Designation | UFV700-230V | UFV700-230V-W | UFV700UL-240V | UFV700UL-120V |
Interior width | 890 mm | 890 mm | 890 mm | 890 mm |
Interior height | 1300 mm | 1300 mm | 1300 mm | 1300 mm |
Interior depth | 605 mm | 605 mm | 605 mm | 605 mm |
Inner doors | 2 | 2 | 2 | 2 |
Unit doors | 1 | 1 | 1 | 1 |
Designation | UFV700-230V | UFV700-230V-W | UFV700UL-240V | UFV700UL-120V |
Sound-pressure level | 47 dB(A) | 47 dB(A) | 47 dB(A) | 47 dB(A) |
Energieverbrauch bei Sollwert -80 °C und Tᵤ = 20 °C | 8.1 kWh/d | 8.1 kWh/d | 8.1 kWh/d | 8.1 kWh/d |
Designation | UFV700-230V | UFV700-230V-W | UFV700UL-240V | UFV700UL-120V |
Number of shelves (std./max.) | 3/11 | 3/11 | 3/11 | 3/11 |
Designation | UFV700-230V | UFV700-230V-W | UFV700UL-240V | UFV700UL-120V |
Freezer racks per compartment | 6 | 6 | 6 | 6 |
528 | 528 | 528 | 528 |
To view available parts for this instrument and instruments like it, click here.
Our parts and accessories inventory is constantly changing. Don't see what you're looking for? Give us a call, we'll be happy to help.
Service Agreements |
Silver |
Gold |
First Class Ship & Fix |
Economy Ship & Fix |
Preventative Maintenance Visit | 1 Preventative Maintenance Visit per Year | 1 Preventative Maintenance Visit per Year | ||
Parts Covered | ||||
Labor Covered | ||||
Unlimited Remote Tech Support | ||||
Depot Repair | ||||
Depot Preventative Maintenance |
GMI was a very professional business to work with. Shipping was quick and easy. Technical Support was quick to respond. I would highly recommend GMI for all your laboratory equipment needs.
Standard Flammable Storage
14 cu. ft. Standard Flammable Storage Freezer with Natural Refrigerants
Standard Flammable Storage
20 cu. ft. Standard Flammable Storage Freezer with Natural Refrigerants