Product Description
High Sensitive Electrochemical Detector for Liquid Chromatography
single cell control, with DC, pulse and scan mode
- Compact
- Thermostat included
With its measurement frequency up to 100 Hz, this electrochemical detector is specially designed for super-fast highly sensitive and selective measurement of oxidisable and reducible substances in (U)HPLC. The AZURA ECD 2.1 comprises of a thermostat-controlled Faraday’s cage, accommodating column and flow cell. The AZURA ECD 2.1 unites the three operating modes DC, Pulse and Scan in one instrument. The DC mode covers about 90% of all applications. The pulse mode is important for PAD (Pulsed Amperometric Detection) of e.g. carbohydrates. The scan mode is used to obtain a voltammogram in method optimization. A digital low-pass filter provides an excellent signal-to-noise ratio. For highest sensitivity, the SenCell flow cell is recommended. The correct flow cell can be chosen from a broad variety of flow cells after our advice.
The AZURA ECD 2.1 is controlled by software packages ClarityChrom® and Chromeleon®
Puls mode
| Measuring range | 10 pA-200 μA in steps of 1, 2 and 5 |
|---|---|
| Filter (cut off) | Advanced Digital Filter, 0.5 -0.001 Hz in steps of 1, 2 and 5 |
| Pulse times | t1: 100-2000 ms; t2: 0-2000 ms; t3: 0-2000 ms in steps of 10 ms |
| Data recording | 20, 40, 60, 80 and 100 ms |
DC mode
| Measuring range | 10 pA-200 μA in steps of 1, 2 and 5 |
|---|---|
| Filter (cut off) | Advanced Digital Filter, 0.5 -0.001 Hz in steps of 1, 2 and 5 |
| Noise | < 2 pA with dummy cell |
Scan mode
| Measuring range | 10 pA-200 μA in steps of 1, 2 and 5 |
|---|---|
| Scan speed | 1-50 mV/s in steps of 1, 2 and 5 |
| Scan cycle | half, complete, continuous |
Detection
| Detector type | Electrochemical detector |
|---|---|
| Version | DC, pulse, scan operating modes |
| Detection channels | 1 |
| Working potential | -2.0 V to +2.0 V |
| Maximum data rate | 10 Hz |
| Autozero | triggered via key, TTL, RS-232 |
Heated zone
| Thermostatting | oven included |
|---|---|
| Temperature range | from 7°C above ambient temperature to 45°C |
Communication
| Control | Parametric control and data-acquisition via LAN port (USB service port) |
|---|
General
| Dimensions | 220 x 440 x 440 mm |
|---|---|
| Weight | 14,6 kg without column and flow cell |





Reviews