Product Description
Thermo Scientific NanoDrop™ Series Spectrophotomers
- Uses patented sample retention technology
- Readings can be produced in less than 5 seconds.
- Sample sizes of as little as 0.5 µL
- Pre-configured methods for common applications such as Nucleic Acid, Protein A280, Microarray, Proteins & Labels, Bradford, BCA, Lowry and Pierce 660nm – and more
- User-friendly software that includes Custom Methods and data export capabilities
- Simple self calibration check using control fluid – no instrument adjustments
- Low-cost operation – no plates or other consumables
The Thermo Scientific NanoDrop™ 2000 features direct easy measurements- Just pipette and wipe. The NanoDrop™ 2000 is much faster and more accurate at measuring using a conventional spectrophotometer. Direct measurement means that accurate readings can be produced in less than 5 seconds.
Tha NanoDrop™ 2000 uses patented sample retention technology that allows for sample sizes of as little as 0.5 µL. The NanoDrop™ 2000 features full spectral output and is capable of measuring DNA, RNA (A260), Protein (A280), concentration and sample purity (260/280 ration). The NanoDrop™ 200 has a concentration range of 2ng/µl through to 15,000 ng/µL dsDNA without dilutions. Other features of the NanoDrop™ 2000 include:
- Pre-configured methods for common applications such as Nucleic Acid, Protein A280, Microarray, Proteins and Labels, Bradford, BCA, Lowry and Pierce 660nm – and more
- User-friendly software that includes Custom Methods and data export capabilities
- Simple self calibration check using control fluid – no instrument adjustments
- Low-cost operation – no plates or other consumables
Other NanoDrop™ Models- NanoDrop™ 2000c, NanoDrop™ 8000 and NanoDrop™ Lite
Thermo Scientific NanoDrop™ 2000c UV-Vis Spectrophotometer with Cuvette Capability
Thermo Scientific NanoDrop™ 2000c– The NanoDrop™ 2000c is a dual-mode UV-Vis Spectrophotometer with Micro-Volume AND
Cuvette Capability for Nucleic Acid and Protein Quantization shares all. The features of the NanoDrop™ 2000c but also features – plus cuvette mode.
Thermo Scientific NanoDrop™ 8000 Multi-Sample Micro-Volume UV-Vis Spectrophotometer
The Thermo Scientific NanoDrop™ 8000 can measure more samples in less time with the same reliability and patented sample retention technology of the NanoDrop™ 2000.
Thermo Scientific NanoDrop™ Lite Basic Microvolume UV-Vis Spectrophotometer
The Thermo Scientific NanoDrop™ Lite is a basic Microvolume UV-Vis Spectrophotometer for simple sample analysis delivered in compact footprint.The NanoDrop™ Lite uses the same sample retention system that has become a hallmark of NanoDrop™ instruments. While it is designed with fewer features than the NanoDrop™ 2000 and NanoDrop™ 8000 series, the NanoDrop™ Lite delivers where it counts: rapid, accurate and reproducible micro volume measurements without the need for dilutions.
NanoDrop™ Pedestal Measurements
The patented sample retention system used by the Thermo Scientific NanoDrop™ 2000 and NanoDrop™ 2000c allows for the analysis of 0.5 µl – 2.0 µl samples, without the need for cuvettes or capillaries.
Nano Drop Model | Lite | 2000 | 2000C | 8000 |
Minimum Sample Size | 1 µl | 0.5 µl | 1 µl | |
Sample Number | 1 | 8 | ||
Path Length | 0.5mm | 1 mm (auto-ranging to 0.05 mm) | 1mm and 2mm | |
Light Source(s) | LED | Xenon flash lamp | ||
Detector Type | Silicon photodiode | 2048 – element linear silicon CCD array | ||
Wavelength Range | 260 and 280 nm | 190 – 840 nm | 220-750nm | |
Wavelength Accuracy | ≤ 8.0 nm | 1 nm | ||
Spectral Resolution | – | < 1.8 nm (FWHM at Hg 253.7 nm) | 3nm (FWHM at HG 546nm) | |
Absorbance Precision | – | 0.002 (1 mm path) | 0.003 absorbance (1 mm path) | |
Absorbance Accuracy | 3% (at 1.05 Abs at 260 nm) | 3% (at 0.74 at 350 nm) | ||
Absorbance Range | 0 – 30 Abs | 0.04 – 300 (10 mm equivalent) | 0.05-75 (10 mm equivalent absorbance) | |
Detection Limit | 4 ng/µl (dsDNA) 0.12 mg/ml (BSA) |
2 ng/µl (dsDNA) | 2.5 ng/µl (dsDNA) | |
Maximum Concentration | 15,000 ng/µl (dsDNA) | 3700 ng/µl (dsDNA) | ||
Measurement Time | < 5 seconds | 20 seconds (for 8 samples) | ||
Footprint | 16 x 11.5 cm | 14 x 20 cm | 24 cm X 32 cm | |
Weight | 0.8 kg | 2.0 kg | 3.4kg | |
Sample Pedestal Material of Construction | 303 stainless steel and quartz fiber | |||
Operating Voltage | 6 vdc | 12 vdc | ||
Operating Power Consumption | 18W | 12 – 18 W (max 30 W) | ||
Standby Power Consumption | ≤ 2.5 | 5 W | 6W | |
Software Compatibility | Local Only | Professional versions of Windows 7 (32 bit and 64 bit), Windows Vista (32 bit), Windows XP (32 bit) | ||
Cuvette specifications for NanoDrop™ 2000c Model Only | ||||
Z-Height | 8.5 mm | |||
Heating | 37 +/- 0.5 °C | |||
Stirrer: | 150-850 rpm | |||
Path Length | 10, 5, 2, 1 mm | |||
Absorbance Range | 0.008 – 1.5 | |||
Detection Limit | 0.4 ng/µl (dsDNA) | |||
Maximum Concentration | 750 ng/µl (dsDNA) | |||
Measurement Time | < 3 seconds | |||
Cuvette Dimensions | 12.5 mm x 12.5 mm, up to 48 mm H | |||
Type | Masked cuvette | |||
UL/CSA and CE | All units are approved to these standards |
Reviews
There are no reviews yet.