Product Description
The Awareness Technology Stat Fax 3300 Microplate Reader is designed to read and calculate the results of in-vitro diagnostic assays. Equipped with a large graphical display, the reader provides absorbance and concentration readings at or near the available wavelength.
The design of this Awareness Technology microplate reader includes many features that reduce user errors and repeats. Among these are stable factory calibration, automatic zeroing, complete operator prompting, and pre-programmed calculations. Detailed labeling, visual and audible feedback, as well as flags and error messages add to the efficient operation of this system.
The flexible Stat Fax 3300 is equipped with multiple, self-prompting operating modes to meet the varying demands and requirements of any facility. The modes include Absorbance Mode, Standard Mode, Rate Mode, Factor Mode, Multi-point Mode, and Index Mode. Moreover, the instrument accepts standard 12 mm round tubes as well as 1 cm square cuvettes.
The Stat Fax 3300 reader‘s user-programmable memory allows the operator to store and name tests in a numbered test menu. Thereafter, these tests may be readily recalled for later use with minimal set up, and remain stored until either changed or deleted by the user. Basic calculations are also permanently stored in the Stat Fax memory, including several single and multi-point equations.
Awareness Technologies Stat Fax 3300 Microplate Reader Specifications
Spectrophotometer Type | Filter photometer |
Optical Configuration | Single beam with continuously rotating filter wheel Monochromatic or bichromatic reading 8 filter positions |
Usable Spectral Range | 330 to 770 nm |
System Procedures | Open and by stored menu |
Calculating Modes | Absorbance, Single Standard, Factor Mode, Multi Standard Mode (up to 7 standards), Multi Standard % Abs (up to 7 standards), Kinetic Mode, Index Mode |
Channels | 120 Open |
Source of Radiation | Tungsten Halogen, 10 Watt, with automatic lamp saver |
Selection of Wavelength | By filter |
Filter Type | 4-cavity interference, long-life ion beam-assisted deposition |
Wavelength Accuracy | +/- 3 nm |
Filter Location | After sample (heat absorbing filter before sample) |
Filter Selection | Automatic by software or via keyboard |
Wavelengths | 340, 405, 505, 545, 580, 630 nm supplied standard, other/additional filters optional |
Half Bandwidth | < 10 nm |
1/100 Bandwidth | 14 nm at 340 nm |
False Radiant Energy Ratio | < 0.001 at 340 and 405 nm |
Cuvette | 1 cm square, 12 cm round, flow through |
Supplied Type | flow through |
Material | 316 stainless, borosilicate windows |
Geometry | Cylindrical, 2.3 mm dia x 5 mm +/- 0.05 mm |
Illuminated Volume | 21 µl |
Minimum Read Volume | 250 µl |
Aspiration/Purge | Vacuum pump at 18 cm of Hg |
Valve | Silicone pinch type |
Cuvette Holder | Thermostatically controlled compartment at 37° C |
Detector | Gallium-Arsenide-Phosphide photodiode |
Display Type | 240 x 128 Graphic LCD w/ Backlight |
Absorbance | -0.5 to 3.5 (flow-through mode) -0.5 to 2.5 (tube or 1 cm cuvette) |
Concentration | Maximum 999,999 |
Kinetic Results | Abs/min with resolution of 0.0002 A/min |
Zero Compensation | Automatic |
Range | -0.5 to 2.5 absorbance |
Parallel | Centronics/IBM-PC compatible |
Serial | RS-232 at 9600 baud, 8 data, 1 stop, no parity Bi-Directional |
Data Input | 1) 20 Key Keypad, 2) PS2 101 Keyboard (connector available at back of instrument) |
Flow-through | < 0.5 % at 1 absorbance, 340/630 nm |
Stability | Better than 0.003 A/hr monochromatic after warm-up Better than 0.001 A/hr bichromatic after warm-up |
Warm-up Time | 90 seconds photometric 15 minutes for temperature compartment |
Electronics | Z180 Microprocessor 18 MHz 128k EEPROM 32 K bytes Nonvolatile RAM (NVRAM) |
Power | Auto-Switching Power Supply Voltage source: 90 – 264 VAC Frequency: 50/60 Hz Power consumption: 60 watts Installation Category: CAT II Fuses: 2.5A/250V Fast 5-20 mm Glass Fuse, 2.0A/250V Fast 5-20 mm Ceramic Fuse, (2) 6/10 250V Slow Blow 3AG Fuses |
Dimensions and Weight | 40cm (L) x 37 cm (W) x 14 cm (H) lid closed (30 cm lid opened), 6.4 kg |