Product Description
- Automated operation
- Responsive touch screen and user-friendly interface
- Language independent icons
- One-step sample processing
- Safely clean probes with automated probe wipe
The Beckman Coulter AcT Diff Hematology Analyzer is an affordable system, ideal for laboratories that carry out up to 50 CBCs per day. Other highlights of the AcT Diff model include a storage capacity of up to 250 patient results, automated sequencing of sample ID number, and automated quality assurance procedures.
The AcT Diff analyzer boasts of a fully automated start-up and shutdown feature to cut preparation time. Instrument operators are also given the option to enter an institutional header onto reports as well as define high/low flagging of patient results
This high performing system is a popular choice among physicians for their office laboratories. Take advantage AcT Diff analyzer‘s innovative features to answer the low-volume needs of small-scale facilities.
Technical Specifications
Methods and Technologies | Coulter Principle counting and sizing Triplicate counting Extend platelet counting Cyanmethemoglobin |
Parameters | WBC, RBC, Hbg, Hct, MCV, MCH, MCHC, Pit, Pct*, RDW, MPV, LY%, LY# Printed WBC, RBC, PLT Histograms *Not intended for diagnostic use |
Sample Analysis Characteristics | Whole blood 12uL, Predilute 20uL, with onboard diluent dispense, Automatic probe wipe, Patient ID with autonumbering, Single button operation, User definable patient sample flagging limits |
Throughtput | Minimum of 50 samples/hour, Results in less than 60 seconds |
Carryover | High to low carryover is less than or equal to 2% |
Printers | Dot Matrix Graphics printer, Single ticket printer, Bubble Jet printer |
User Interface | Touch screen, Language independent icons, Single screen display of all results, One button sample aspiration |
Host Communication | RS232 with ASTM standards |
Dimensions | |
Height | 43cm (16.5 inches) |
Width | 38cm (15 inches) |
>Depth</b | 39cm (15.4 inches) |
Weight | 14 kg (30 lbs) |
Power/Input | |
100 plus/minus 10% | 50/60 Hz |
120 plus/minus 10% | 50/60 Hz |
220 plus/minus 10% | 50/60 Hz |
240 plus/minus 10% | 50/60 Hz |
Linearity | |
WBC x 1000 cells/uL | Linearity Range: 0 to 99.9 Limits (whichever is greater): +/- 0.3 or +/- 5.0% Linearity Range: 0 to 7.0 |
RBC x 1,000,000 cells/uL | Linearity Range: 0 to 7.0 Limits (whichever is greater): +/- 0.04 or +/- 2.0% Linearity Range: 0 to 25.0 |
Hgb g/dL | Limits (whichever is greater): +/- 0.2 or +/- 3.0% Linearity Range: 0 to 999 |
Plt x 1000 cells/uL | Limits (whichever is greater): +/- 10 or +/- 10.0% |
Background Counts | WBC +/- 0.4 x 1000 cells/uL RBC +/- 0.04 x 1,000,000 cells/uL Hgb +/- 0.2 g/dL Plt +/- 7.0 x 1000 x 1000 cells/uL |
Precision | |
WBC at 6.0 – 15.0 x 1000 cells/uL | CV: Less than 3.0% |
RBC at 3.00 – 6.00 x 1,000,000 cells/uL | CV: Less than 3.0% |
Hgb at 12.0 – 18.0 g/dL | CV: Less than 2.0% |
MCV at 80.0 – 100.0 fL | CV: Less than 3.0% |
RDW at 12.0 – 15.0% | CV: Less than .0% |
Plt at 200 – 500 x 1000 cells/uL | CV: Less than 7.0% |
MPV at 5.0 – 20.0 fL | CV: Less than 3.0% |
LY at 20 – 50 % | CV: Less than 1.5% (S.D.) |
MO at 2.0 – 10% | CV: Less than 1.5% (S.D.) |
GR at 20.0 – 70.0 % | CV: Less than 3.0% (S.D.) |
Reviews