Beckman LS 6500 Scintillation Counter
Beckman LS 6500 Scintillation Counter
| Product Description |
The Beckman LS 6500 Liquid Scintillation Counter is a high performing system that offers unparalleled flexibility. Beckman Coulter combined technological advancements, accuracy and configurability to produce this reliable system.
Beckman LS 6500 systems are equipped with a 32,678 channel analyzer that delivers an effective resolution of 0.06 keV for accurate results across a variety of samples. A patented H-Number Plus external quench monitor (H-Number Plus) fitted into the system provides accurate Defects Per Million (DPM) calculations and Automatic Quench Compensation. Both measurements are independent of sample volume or cocktail composition. The feature is also the basis for instrument calibration, color detection and correction, two-phase monitoring, unquenched window determination and multi-label DPM quench curves.
Instead of lowering the compartment to the sample chamber, the LS 6500 counter raises the sample plastic racks to the sample chamber. This design prevents possible contamination of the chamber with dust or liquid residue. As for system operations, the combination of easy-to-use menus, context-sensitive help screens and 50 user programs makes the LS 6500 a one-button system.
The System Plus and System Advantage models are equipped with powerful analysis tools. Units come with a single or dual label DPM, an exclusive H-Number Plus quench monitor as well as the Versa-Rack™ vial system which allows mixing of 20 mL and 7 mL vials. Enhancement packs and other features can be added later to address emerging laboratory needs.
Call us today to learn more about the Beckman LS 6500 Liquid Scintillation Counter!
| Beckman LS 6500 | |
| Resolution | 32,768 channels, 0.06 keV per channel |
| Reproducibility | +/- 1% Counting Efficiency |
| Maximum Count Rate | Total Singles: 26 x 10^6 CPM 3H Window: 10 x 10^6 CPM 14C Window: 13 x ^6 CPM Above 14 C Window: 5 x ^6 CPM |
| Power Requirements | 120 / 240V / BTU HR with 50/60 Hz |
| Instrument | 3A / 1.5A / 1230 |
| Temperature | 7.5A / 3.8A / 3060 |
| Inrush Current | 5A at 120V, 2.5A at 240V |
| Power Failure Recovery | System returns to a disrupted Automatic Count with fully charged battery |
| Dimensions (W x H x D / Weight) | Instrument: 91.4 x 67.3 x 80 cm / 210 kg Refrigeration Acc.: 35.5 x 56 x 73.7 cm / 45.4 kg Printer: 39 x 10 x 30.5 cm / 5 kg |
| Ambient Temperature Range | 15 - 35'C |
| Maximum Relative Humidity | 85% |
| Ambient Temperature Range | 15'C to 30'C |
| Maximum Pull Down Time | 8 Hours |
| Set Point 1 | 12'C +3'C |
| Set Point 2 | 15'C +3'C |
| Set Point 3 | 18'C +3'C |
| Efficiency | 2H: >/= 60% in a wide window 14C: >/= 95/5 in a wide window |