You are here: GMI Clinical Clinical Blood Gas Analyzer YSI 2300 STAT Plus Glucose & Lactate Analyzer

New and Used Lab Equipment - GMI Inc

Loading
Clinical arrow Blood Gas Analyzer arrow YSI 2300 STAT Plus Glucose & Lactate Analyzer (WS-2300STAT+)

YSI 2300 STAT Plus Glucose & Lactate Analyzer

YSI 2300 STAT Plus Glucose & Lactate Analyzer

  • Results in one minute or less
  • Proven Technology
  • Precise, accurate results
  • Whole-blood capability
  • Small sample size (25 uL)
... Full Description >>>


Complete Product Description

The YSI 2300 STAT Plus is the only dedicated glucose and lactate analyzer available. This analyzer makes precise glucose measurements in whole blood, plasma or serum; precise lactate measurements in whole blood, plasma or cerebrospinal fluid. The analyzer's hematocrit correction option allows you to have glucose results expressed as plasma even when running whole blood.

YSI 2300 STAT Plus Features

Speed

  • Results in one minute or less

Low cost per test

  • The immobilized enzymes do not have to be renewed after each test, minimizing the use of costly reagents and supplies

Proven Technology

  • You can be confident that results are accurate

Precise, accurate results

  • YSI provides the highest precision and most accurate results for whole-blood glucose and lactate.

Whole-blood capability

  • The YSI 2300 is the only dual-channel glucose and lactate analyzer that uses whole blood, so you do not waste spinning blood.

Small sample size (25 uL)

  • The analyzer requires a small sample, making it practical in neonate applications

Simple, safe operation

  • Menu-driven software guides operators through each step. Diagnostics provide performance analysis and troubleshooting. Samples are introduced to the analyzer by tubes or syringe; no needles required.

Batch operation

  • The 24-position turntable lets you run batches. You can assign each batch and sample an identification number with date and time.

Autocalibration

  • Calibration is automatic, so the analyzer is always ready.

Interference rejection

  • YSI immobilized enzyme technology eliminates most of the interferences that plague other methods.

Broad measurement range

  • Up to 900 mg/dL glucose and 30 mmol/L lactate virtually eliminates the need for sample dilution.

YSI 2300 STAT Plus Specifications

Sample Size
25 uL (aspirated volume)
Respose Time (from test tube holder)
Normal mode Displayed and printed results in 65 seconds
  Sample to sample interval is 100 seconds.
Screen Mode Displayed and printed result in 45 seconds.
Measurement Range
Normal mode Glucose: to 900 mg/dL
(9000 mg/L, 50.0 mmol/L)
Lactate: to 30.0 mmol/L
(267 mg/dL, 2670 mg/L)
Screen Mode Glucose: to 500 mg/dL
(5000 mg/L, 27.8 mmol/L)
Lactate: to 15.0 mmol/L
(134 mg/dL, 1335 mg/L)
Precision
Glucose Larger of 2% of reading or 2.5 mg/dL
(25 mg/L or 0.2 mmol/L)
Lactate Larger of 2% of reading or 0.1 mmol/L
(1 mg/dL or 10 mg/L)
Calibration Point
Glucose Larger of 2% of reading or 2.5 mg/dL
(25 mg/L or 0.2 mmol/L)
Lactate Larger of 2% of reading or 0.1 mmol/L
(1 mg/dL or 10 mg/L)
Calibration Point
Glucose 180 mg/dL (1800 mg/L, 10.0 mmol/L)
Lactate 5.00 mmol/L (45 mg/dL, 445 mg/L)
Membrane Working Life (typical)
Glucose 21 days
Lactate 14 days
Operating Temperature
15 to 35C
Humidity
10 to 90% noncondensing
Power
110-120 VAC or 220-240 VAC, 50-60 Hz
Size
25.4 x 35.6 x 35.6 cm, 11.4 kg
10 x 14 x 14 inches, 25 lbs
Regulatory Compliance
CE, FDA Registered Class II (862.1345), CSA
CE Device Category
IVD