COULTER JT Series
Hematology Systems
 |
Diluter Keypad: Push-button control of all diluter functions
including start-up and shutdown. |
 |
Bar code reader and closed-vial sample port: Simply pass
closed vial sample with bar code label past reader and place tube in instrument.
|
 |
Secondary Mode: Convenient whole blood microsample mode for
low volume aspiration. |
 |
Data Terminal: System control center for processing,
display, storage and transmission of data. |
|
JT
Parameters
- 16, including Coulter Histogram Differential
Interpretive Report
Throughput
Closed-Vial Sampling
Sample Volume
- Closed-Vial Mode: N/A
- Manual Whole Blood Mode: 110 µL
Bar Code Sample ID
Data Storage
- Results with Histogram: 270/37
Quality Control
- 9 Files: Yes
- Levey-Jennings Graphs: Yes
- XB: Yes
|
JT
Parameters
- 16, including Coulter Histogram Differential
Interpretive Report
Throughput
Closed-Vial Sampling
Sample Volume
- Closed-Vial Mode: 270 µL
- Manual Whole Blood Mode: 110 µL
Bar Code Sample ID
Data Storage
- Results with Histogram: 270/37
Quality Control
- 9 Files: Yes
- Levey-Jennings Graphs: Yes
- XB: Yes
|
JT3
Parameters
- 18, including Coulter Histogram Differential
Interpretive Report
Throughput
Closed-Vial Sampling
Sample Volume
- Closed-Vial Mode: 270 µL
- Manual Whole Blood Mode: 110 µL
Bar Code Sample ID
Data Storage
- Results with Histogram: 2000/280
Quality Control
- 9 Files: Yes
- Levey-Jennings Graphs: Yes
- XB: Yes
|
Input:
 |
115 +/- 10 Vac, 60 Hz
|
 |
100 +/- 10 Vac, 50/60 Hz
|
 |
200 +/- 10 Vac, 50/60 Hz
|
Consumption:
460 W (1,571 BTU/hr.)
Precision (n greater than or
equal to 31)
WBC at 10.0 x 1000 cells/µL
 |
CV: Less than 2.0%
|
RBC at 5.00 x 1,000,000 cells/µL
 |
CV: Less than 2.0%
|
Hgb at 15.0 g/dL
 |
CV: Less than 1.0%
|
MCV at 90.0 fL
 |
CV: Less than 2.0%
|
RDW at 13.0%
 |
CV: Less than 2.2%
|
Plt at 300.0 x 1000 cells/µL
 |
CV: Less than 4.0%
|
Plt at 30.0 x 1000 cells/µL
 |
CV: Less than 8.0%
|
Plt at 10.0 x 1000 cells/µL
 |
CV: Less than 12.0%
|
MPV at 9.0 fL
 |
CV: Less than 2.2%
|
Lymph # at 0.6 x 1000 cells/µL
 |
CV: Less than 5.0%
|
Mono # at 0.2 x 1000 cells/µL
 |
CV: Less than 10.0%
|
Gran # at 0.6 x 1000 cells/µL
 |
CV: Less than 5.0%
|
|
|
|
|
|
|
|
COULTER JT Series
Hematology Systems
Carryover
Less than or equal to 2%.
Throughput
Up to 60 samples/hr
Linearity
WBC x 1000 cells/µL
 |
Linearity Range: 0 to 99.9
|
 |
Limits (whichever is greater):
+/- 0.3 or +/- 3.0% |
RBC x 1,000,000 cells/µL
 |
Linearity Range: 0 to 7.0
|
 |
Limits (whichever is greater):
+/- 0.04 or +/- 2.0% |
Hgb g/dL
 |
Linearity Range: 0 to 25.0
|
 |
Limits (whichever is greater):
+/- 0.2 or +/- 2.0% (< to
4% from
18-25 g/dL in closed vial mode) |
MCV fL
 |
Linearity Range: 50 to 200
|
 |
Limits (whichever is greater):
+/- 3.0 or +/- 3.0% |
Plt x 1000 cells/µL
 |
Linearity Range: 0 to 999
|
 |
Limits (whichever is greater):
+/- 10 or +/- 7.0% |
MPV fL
 |
Linearity Range: 5.0 to 20.0
|
 |
Limits (whichever is greater):
+/- 1.0 or +/- 3.0% |
Dimensions
Analyzer
 |
Height: 51.5 cm (20.25")
|
 |
Width: 77.5 cm (30.5")
|
 |
Depth: 58.4 cm (23.0")
|
 |
Weight: 51.3 Kg (113 lb)
|
Data Terminal
 |
Height: 41 cm (16")
|
 |
Width: 48.5 cm (19")
|
 |
Depth: 54 cm (21.25")
|
 |
Weight: 21.3 Kg (47 lb)
|
Mode to mode comparison
 |
WBC +/- 0.4 x 1000
cells/µL or 5%, whichever is greater |
 |
RBC +/- 0.2 x 1,000,000
cells/µL or 2%, whichever is
greater |
 |
Hgb +/- 0.3 g/dL or 2%,
whichever is greater |
 |
Plt +/- 20 x 1000
cells/µL or 7%, whichever is greater |
When measured in accordance with the
Product Reference
Manual.
Electronic Stability
Change in calibration of the electronic measurement system
is less than 1% per month when measured in accordance with the Product Reference Manual
and compared at monthly intervals.
Sample Volume
110 µL capillary or venous whole blood -
secondary mode
270 µL venous whole blood -
primary (closed vial) mode
COULTER JR Series will operate with the majority of
collection devices available worldwide, ranging in size from the 2 mL through the 7 mL
tubes.
Catalog Listings
The COULTER JR Series includes Main Unit Analyzer, Data
Terminal, Data Storage Package and Printer. Choose a Single Ticket or Graphic printer to
report your results.
For maximum efficiency Coulter's Auto Reporter printer is
available as an option. This printer collates bar code labeled tickets and sorts normal
and abnormal reports.
Data storage on the JT3 Data Terminal provides memory
capacity for 2000 patient CBC profiles or 280 numeric results with histograms (the JT and
JT2 models store 270 patient profiles or 37 composite results with histograms. Optional
additional storage data is available).
One Data Storage package can be added to increase memory
capacity by 1800 numeric profiles or 250 composite results with histograms for a total of
over 3900 CBC profiles or 540 reports with histograms.