Introducing the Hettich ROTIXA 50S. Among experts Hettich ROTIXA has become a household word for high sample throughput. This is why ROTIXA models are the centrifuges
of choice in development and quality control laboratories of the life science industry and in medical routine laboratories.
With generating a maximum RCF of 5-503 ROTIXA centrifuges are suited for separating a wide range of sample materials such as samples
of food- cell suspensions- sediments of chemical reactions- soil and- of course- blood and blood components.
In addition to all commonly used blood collection systems the Hettich ROTIXA 50S centrifuges accommodate Falcon tubes- microtitre
plates and centrifuge bottles up to a volume of 1-000 ml. With the growing use of clinical chemistry analysers in mind- the ROTIXA 50S centrifuges are also designed to spin racks of well-known manufacturers like Olympus- Hitachi- LKB- and Behring. It goes without saying that they also accept the time-tested round-bottom tubes up to a max. volume of 250 ml.
The controls and displays of the ROTIXA 50S centrifuges are arranged according to ergonomical aspects. The large display
is clearly arranged and well readable. During centrifugation the actual values of all parameters are indicated. Entry of the parameters
is exact- fast and easy with selector keys on the water-protected foil keypad and an adjusting knob.
Speed n- RPM input in increments of 10
Relative centrifugal force RCF input in increments of 1
Radius r/mm needs to be entered when speed or RCF have to be determined
Running time t input in minutes and seconds- max. 999 min.
Run-up/run-down time input of acceleration and braking ramps in levels 1 - 9- or timecontrolled in minutes and seconds- 0=unbraked run-down.
Deceleration with brakeforce cut-off input of the cut-off speed in RPM
Integral of the centrifugal acceleration provides continuous information on the accumulated RCF the centrifuged material is exposed to during centrifugation.
Temperature T/C with the ROTIXA 50 RS input in increments of 1 C within a range of ? 20 C to + 40 C
Memory Since saving a lot of time the storing of programmes is indispensable in laboratories with many
different and complicated centrifuging tasks. Up to 89 complete running programmes can be memorized and recalled with their storing number. They remain in the memory even after the centrifuge has been switched off.
For special routine or scientific applications the ROTIXA 50 S centrifuges can be fitted with the following optional equipment:
Centrifuging data report
With the Hettich data report system
up to 29 centrifuges can be
controlled by a single PC.
An optical cable connects the centrifuge
to a BUS system. Protocols of
selected centrifuging data and barcode
readers can be produced via a
PC with an RS 232 interface.
Visual indication when rotor is
stationary
for additional information on the
completion of a centrifuging
cycle.
Chaining of programmes
allows to programme the centrifuge
to run different programmes
in succession.
Second independent tacho
system
for calibrating the speed measuring
instrument as required by
quality management systems.
Whether entering
single parameters
or chaining
complete running
programmes- the
clearly arranged display
with its selector
keys and adjusting
knob provides
for extreme
user-friendliness.
| Specifcation | |
| Performance | 4-900 min-1 in a swing- out rotor corresponding to an RCF of 5-503- max. capacity of 4 x 1-000 ml. |
| Extremely short run -up and rundown times. |
|
| General | Smooth and quiet running. |
| Frequency-controlled drive- maintenance only required according to UVV regulations. |
|
| See- through glass in the lid for additional speed control. |
|
| Saftey | Lid locking and holding |
Lid dropping protection |
|
| Centrifuging chamber of stainless steel |
|
| Key-operated switch for locking and protecting programmes |
|
| Motor overheating protection | |
| Chamber overheating protection in the ROTIXA 50 RS |
|
| Imbalance switch-off | |
| Automatic rotor recognition | |
| Conformity with the latest national and international safety standards such as IEC 1010-2-2- EN 61010-1 and all relevant EC regulations |
|
| Capacity | |
| General | 120 x15 ml |
| Falcon Tubes | 32 x 50 ml |
| Microtitre plates | 16 Standard Size |
| HITACHI Racks | 20 |
| Bottles/ Blood Bags | 4x 1000 ml/ 4 bags |
| Technical | |
| Power Supply | 208-240 V1~ |
| Frequency | 50-60 hz |
| Consumption | 2.7 kVA |
| Radio Interference | EN 55011 ISM class B |
| Immunity | EN 50082-2 |
| Range | |
| Capacity | 4x100ml (swing out rotor) |
| Speed /RCF | n= 4900 min /RCF 5503 |
| Centrifuging Chamber Temp | controllable -20C to 40C |
| Running Time | 1-999 min (continuous run) |
| Dimmensions | |
| 50S Model H x W x D inch | 41.2 x 26 x 28.8 |
| Weight | 254 Ilbs |