Beckman Coulter AcT
Diff
Hematology Analyzer
Methods and
Technologies:
Coulter Principle counting and sizing
Triplicate counting
Extended platelet counting
Cyanmethemoglobin
Parameters:
WBC, RBC, Hbg, Hct, MCV, MCH, MCHC, Pit, Pct*, RDW, MPV, LY%,
LY#, MO%, MO#, GR%, GR#, PDW* Printed WBC, RBC, PLT Histograms *Not
intended for diagnostic use.
Sample Analysis Characteristics:
Whole blood 12µL,
Predilute 20µL, with onboard diluent
dispense, Automatic probe wipe, Patient ID with autonumbering,
Single button operation, User definable patient sample flagging
limits
Throughtput:
Minimum of 50 samples/ hour,
Results in less than 60 seconds
Carryover:
High to low carryover is less than or equal to 2%
Printers:
Dot Matrix Graphics printer, Single ticket printer, Bubble Jet
printer
User Interface:
Touch screen, Language independent icons, Single screen display
of all results, One button sample aspiration
Host
Communication:
RS232 with ASTM standards
Dimensions/Weight:
Height: 43cm (16.5 inches)
Width: 38cm (15 inches)
Depth: 39cm (15.4 inches)
Weight: 14 kg (30 lb)
Power/Input:
100 plus/minus 10% 50/60Hz
120 plus/minus 10% 50/60Hz
220 plus/minus 10% 50/60Hz
240 plus/minus 10% 50/60Hz
Linearity:
WBC x 1000 cells/µL
 |
Linearity Range: 0 to 99.9
|
 |
Limits (whichever is greater):
+/- 0.3 or +/- 5.0%
|
RBC x 1,000,000 cells/µL
 |
Linearity Range: 0 to 7.0
|
 |
Limits (whichever is greater):
+/- 0.04 or +/- 2.0%
|
Hgb g/dL
 |
Linearity Range: 0 to 25.0
|
 |
Limits (whichever is greater):
+/- 0.2 or +/- 3.0%
|
Plt x 1000 cells/µL
 |
Linearity Range: 0 to 999
|
 |
Limits (whichever is greater):
+/- 10 or +/- 10.0%
|
Background Counts:
 |
WBC +/- 0.4 x 1000
cells/µL |
 |
RBC +/- 0.04 x 1,000,000
cells/µL |
 |
Hgb +/- 0.2 g/dL
|
 |
Plt +/-
7.0 x 1000 x 1000
cells/µL |
Precision:
WBC at
6.0-15.0 x 1000 cells/µL
 |
CV: Less than
3.0%
|
RBC at
3.00-6.00 x 1,000,000 cells/µL
 |
CV: Less than
3.0%
|
Hgb at
12.0-18.0 g/dL
 |
CV: Less than
2.0%
|
MCV at
80.0-100.0 fL
 |
CV: Less than
3.0%
|
RDW at
12.0-15.0%
 |
CV: Less than
.0%
|
Plt at
200-500 x 1000 cells/µL
 |
CV: Less than
7.0%
|
MPV at
5.0-20.0 fL
 |
CV: Less than
3.0%
|
LY
at 20-50%
 |
CV: Less than
1.5% (S.D.) |
MO at
2.0-10.0%
 |
CV: Less than
1.5% (S.D.)
|
GR at
20.0-70.0%
 |
CV: Less than
3.0% (S.D.) |
You can reserve/order this
instrument by calling
763-712-8717 or by
completing the form
on the Ordering Page

|