|
377 SEQUENCING GEL
PREPARATION
Introduction
This SOP describes how to set up plates, and prepare a gel for a sequencing
377 run.
Materials
10% Ammonium Persulfate (see
SOP/HH5TA02/1998/0046)
TEMED
Accugel
Glass beaker
25 ml stripette and electronic pipetter.
377 glas plate pair [Ask Jackie, these are specially etched]
377 0.2 mm spacers
377 0.2 mm sharks tooth well former comb
Black Cassette
Method
For 1 Gel - Accugel
30 ml acrylamide
150 µl 10% APS
21 µl TEMED
Preparing the plates
* The glass plates for the 377 machines are
long plates with small indents at the base of the back plate, and notches
at the top of the front
plate
* The plates are produced in pairs and must stay within these pairs, each
plate also has an inside and an outside.
* The outside can be identified as all the plates are marked with either a
manufacturers stamp or with an identification number written with a
diamond pen, some may even had OUT written on the outside.
* The spacers for the 377 gels are long with a notch at the top. This
notch faces the inside of the gel.
1) Place the bottom plate in cassette, making sure that the notched ends are
against the silver locating pins. Wipe the spacers before use, then place
them down the long side of the back plate, then put the top plate on and
clip all but the top clips in place.
2) Slide the black clamp onto the cassette
and under the glass plates. Tighten the white plastic screws so that the
black clamp bends a little at each end. [it is important to tighten these
screws as otherwise the comb will not be tight in the gel and samples will
leak.
3) Make up the gel mix as above, in clean
glassware, using the electronic pipette to mix it by pipetting slowly up and
down, taking care not to introduce bubbles. Add the gel mix slowly with an
electronic pipette to the top of the gel, tapping on the top of the glass
plate at the gel front to stop bubble formation.
4) Once the gel has filled the plate insert
the comb. These are found in the racks next to the sink. The flat edge of
the sharks tooth is placed into the gel with the teeth sticking out the top.
5) Clamp the gel by the comb with the
transparent clamp and clip the two clips on the cassette.
6) Allow to set for 2 hours before use. If
the gel is to be left for a long period before use it must be wrapped, top
and bottom, with Kimwipes soaked in 1xTBE, then cling film.
7) The gels are numbered sequentially to
allow tracking in the event of any problems and logged into the gel book in
the gel pouring room.
Safety Issues
Unpolymerised acrylamide is a powerful
neurotoxin and putative carcinogen. It should always be handled with gloves
and eye protection must be worn to prevent ingress by splashing to eye.
Refer to COSHH form WTCHG97-24. Polymerised acrylamide cannot traverse skin
but gel matrix always contains small amounts of monomer so gloves should be
worn whenever handling gels.
USE OF
AN ABI 377 FOR SEQUENCING
Introduction
This SOP describes how
to use an ABI 377 for sequencing.
Materials
377 sequencing gel
(SOP/QH58Z02/1999/0062)
1xTBE (SOP/QH58Z02/1999/0027)
Sequencing Samples
Method
1) Preparing the machine
After the previous run is completed remove
the heating plate and place handle side down on the bench [DO NOT PLACE
FLAT SIDE DOWN AS THIS CAN RUIN THE HEATING PLATE]. Remove the used
cassette and take it complete with the upper buffer chamber to the
sink.Take off upper buffer chamber and wash with RO water before use. Take
used gel out of the cassette and place in tray provided ensuring that
there is tissue between the top of the gel already in the tray and the gel
to be put in. This prevents scratching of the glass plates. Wash the
cassette with RO water and dry before use.
Discard buffer from the lower buffer chamber and wash with RO water.
Collect a sequencing gel from the gel pouring room , wash off excess
acrylamide with warm water followed by deionised water and dry
plates using white kimwipe, lint free tissues (red box).
Take out the pouring comb and make sure there is no acrylamide left
in-between the gel plates at the top.
Clean between the plates with an old loading comb to remove any of the
old acrylamide.
Wash the well left by the sequencing comb with 1 x TBE and carefully
insert the sharks tooth comb into the gel so that the tips of the teeth
just penetrate the top of the gel.
Do not push the comb into the gel too far otherwise it will be difficult
to load the gel.
If the comb has been pushed into the gel too far but is still loadable DO
NOT remove the comb and try again as this will ruin the wells.
Clip the gel into the ABI cassette ensuring the gel plates are flat in the
cassette.
2) Preparing the computer
Before each run, restart the Mac computer
by going to the Special menu and selecting Restart. This
ensures that the RAM memory is not fragmented and will restart the
collection software.
Select ABI Collection Software, this nornmally comes up automatically when
the computer is restarted. In the File menu select New then
select Sequencing Run.
In the Run window that appears make sure
that the following settings are present.
For dRhodamine and BigDye Terminators.
PLATE CHECK: Plate Check A.
PRE RUN: Seq PR 36E-2400
RUN: SEQ RUN 36E-2400.
MATRIX: ABIX_BDT_ACCUGEL_DEC99
For the matrix setting the X refers to the
machine so ABI A will have a matrix called ABIA_BDT_ACCUGEL_DEC99
[Note: On the Monaco ABI the pre run and
run modules have the suffix chiller added because of the additional water
bath installed]
To change the number of lanes run select
either 36 and full scan or 48 and XL scan.
Plate Check
Press plate check and then the scan window will appear and after
approximately 1 min 4 coloured lines will appear. These should be flat
and below 2048. If there are peaks in this trace this indicates that the
plates are not clean and the cassette needs to be taken out of the machine
and cleaned with kimwipes again. If the peak is not removed the
corresponding lane may have to be missed out during loading.
If the plate check is OK
Clip on the buffer chamber and fill with 1
x TBE. Wait a minute to ensure there is no leakage of buffer down the
front of the plate. This could cause arching of electrical current if left
and severely damage the ABI.
Next clip on the heating plate and fill the lower buffer chamber with 1 x
TBE.
Rinse out the wells thoroughly to ensure that there are no air bubbles or
excess acrylamide present.
Pre Run
Select pre-run. Then go to Window: Status and check the run is for
1 hour and that the gel gets up to temperature 51oC.
You should allow at least 10 minutes for the pre-run.
While the gel is pre running prepare the sample sheet by selecting File
- New -Sequencing Sample sheet and typing in the appropriate lanes.
Save the sample sheet.
Prior to loading your samples and performing a run, add 3 µl loading dye
for ABI377. Make sure the pellet, which is not visible, is thoroughly
resuspended either by pipetting the loading dye up and down the tube a few
times or by vortexing gently for 10-15 seconds to ensure that your sample
is resuspended in the loading dye. Centrifuge samples briefly.
Denature the samples for 2 mins @ 95 oC
and immediately store on ice until ready to load (this is important as DNA
will re-anneal if not immediately placed on ice).
Pause the Pre Run and wash out the wells once more.
Load sample sheet into the run window [ the machine will not be able to
start a run without a sample sheet]
Load 1-2 µl per lane, loading every other lane and then cancel the pause-
press continue - and run the samples in for 3 minutes.
Pause the gel again and wash out the wells and load the well not loaded in
the first round. Run the samples in for a further 3 minutes and then
cancel the Pre Run. Start the Run and then a window will
appear asking to name the run.
After the run has finished the run folder can be transferred to another
computer for further analysis
[The reason for loading every other lane is
that the machine can recognize the individual lanes and track the gel in
the analysis program. If every lane is loaded at the same time with a
sharks tooth comb the machine will only see one broad lane]
PREPARATION OF A
SEQUENCING SAMPLE SHEET
Introduction
This SOP describes how to produce a sample sheet for sequencing.
Materials
List of samples.
Method
373 e.g. ABI 11 In the File menu select New Sample sheet. Fill in the
Sample file name then the Dye Set/Primer name. For BigDyes this is "373
BDT" and for dRhodamine this is "373 dRDT". There are 4 choices which can
be selected under the heading Auto Settings ? AA = auto analyse, CS =
contains sample, PR = plot raw data, PA = plot analysed data.
373-XL and 377 e.g. ABI 10 In the Data
Collection Software select File then New then Sequence Sample. Fill in
the Sample name then the Dye Set/Primer name. For BigDyes this is "373
BDT" and for dRhodamine this is "373 dRDT". Select the appropriate
matrix, this must be done if automatic analysis has been selected.
Safety Issues
None.
PRODUCTION OF SEQUENCING MATRIX
USING DATA UTILITY
Introduction
This S.O.P. describes the production of a sequencing matrix on both 373 and
377 ABI machines.
Materials
373 or 377 gel (see SOP/HH5TA02/1998/0023
and SOP/HH5TA02/1998/0010)
BigDye dRhodamine matrix dye set
Deionised Formamide
Method
Set up the gel in ABI machine as
SOP/QH58Z02/1999/0066, SOP/QH58Z02/1999/0067 or SOP/HH5TA02/1998/0012, for
ABI373, ABI 373-XL and ABI 377 respectively.
Combine 2.5 µl of each matrix
standard with 2.5 µl deionised formamide, in duplicate. This will give
you 8 samples, 2 of each dye type. Heat each sample to 96oC
for 2 minutes to denature before loading.
Create a sample sheet to assign sample (see
SOP/QH58Z02/1999/0064) but do not assign a matrix to the samples and do
not allow autoanalysis.
Load the Matrix Standards in duplicate, 2.5
µl of the matrix standards, i.e. load 8 lanes 2 dR110, 2 dR6G, 2 dROX and
2 dTAMRA, leaving an empty lane between each standard. Select the
appropriate filter-set, and start the run. 377 ? Run modules containing
filter set E, 373 ? filter-set A.
After the gel has run
Open up Sequence Analysis software
Go to File, then Open, then Collection gel and open
the gel you wish to produce a matrix from.
Track the lanes containing the Matrix
standards - this is done by selecting a lane and going to Gel on
the menu and selecting Mark lane used then move the tracking lanes
so that they follow exactly the lane of that well. When this has been
done for all the lanes used, go to Gel and Extract the
lanes.
From within the Utilities folder of the
Sequence analysis software, open the DataUtility program. In the
Utilities menu, choose Make Matrix.
In the dialogue box that appears make sure that the Dye Primer matrix
button is selected. In the next box that appears press the "C" nucleotide
button and select the appropriate sample file for that nucleotide
according to the table below.
Box Dye Primer Matrix Taq Terminator Matrix T7 Terminator matrix
| C |
dR110 |
dROX |
dRG6 |
| A |
dRG6 |
dRG6 |
dTAMRA |
| G |
dTAMRA |
dR110 |
dROX |
| T |
dROX |
dTAMRA |
dR110 |
For each sample fill in a start point, where
analysis of each sample will start, start with the default value of 2000 and
for the points choose 1500, for the number of data points to analyse.
Click New File and name the new matrix
e.g. BigDye ABI10 Oct99.
Then press OK.
The computer will then try and make the
matrix, if it is unable to do this an error box will appear telling at which
point it has failed. One solution if it has failed to make a matrix is to
increase the amount of data points to analyse.
If the matrix has been successful go back to
the Utilities menu, choose Make Matrix. This time make sure
that the Taq Terminator button has been selected before choosing the sample
files. After selecting the sample files, select Update the File and
choose the correct matrix name e.g. BigDye ABI10 Oct99 to update.
Repeat again as above but this time making
sure that the T7 Terminator Matrix button has been selected.
The matrix is automatically saved in the ABI
folder.
To check the matrix, go to the utilities menu
and select copy matrix. Under Source, select Instrument file and choose the
matrix you wish to check. The 3 matrix files are all shown at once and
should look like the box below.
|
1.000
|
0.005
|
0.092
|
0.168
|
|
0.643
|
1.000
|
0.055
|
0.024
|
|
0.012
|
0.564
|
1.000
|
0.355
|
|
0.022
|
0.010
|
0.444
|
1.000
|
|