Awareness Technologies Stat Fax 2100 Microplate Reader
Awareness Technologies Stat Fax 2100 Microplate Reader
| Product Description |
The Awareness Technology Stat Fax 2100 Microplate Reader is a compact, fully automatic reading unit. With its biochromatic optics, the reader is most ideal for in-vitro diagnostic use, computing the results of endpoint colorimetric assays.
A versatile unit, the Stat Fax 2100 can measure monochromatically or bichromatically. The reader also accepts many single column microstrips up to 12 wells in length, as well as break-apart wells in both flat- and round bottoms. After positioning the carrier to the desired row, the system automatically positions the strips, reads absorbances, calculates concentrations, and prints results. The integrated thermal graphics printer also allows the system to plot standard curves.
Additionally, the Stat Fax 2100 instrument offers pre-programmed modes for performing the most commonly used calculations, multiple convenience features, and user programmable memory to further facilitate testing. Each mode comes with a step-by-step prompting, reducing error and streamlining operation for the user. User-programmable memory also allows the operator to store test protocols, recall tests by number, and delete unwanted tests from a user-generated test menu.
The convenient Stat Fax 2100 reader combines automatic blanking options with the location indication of positive and negative controls. Moreover, the user can enter control acceptance criteria for automatic comparisons, and edit out discrepant duplicates with automatic reread and recalculation.
Call us today to learn more about the Awareness Technology Stat Fax 2100 Microplate Reader!
| PHOTOMETRIC | |
| Linear Measurement Range | 0.00 to 3.0 Absorbance Units (A) |
| Photometric Accuracy | ± (1% of the reading + 0.01 A) from 0 to 1.5A ± (2% of the reading + 0.01 A) from 1.5 to 3.0A |
| Stability | Drift of no more than 0005A in 8 hours |
| Warm up Time | 45 seconds, built-in |
| Light Source | Tungsten Lamp |
| Standard Wavelengths | Standard: 405, 450, 492, 630nm Six Filter UV: 405, 450, 492, 630, 545, and 340 nm Six Filter VIS: 405, 450, 492, 630, 600, and 545 nm |
| Type of Filter | Multiple cavity sealed interference, 10 nm half bandpass typical |
| Vessel | Standard 96 well plates (round or flat bottomwells); also accepts strip trays of standard size (8x12 configuration) |
| ELECTRONIC & SOFTWARE | |
| Microprocessor | Z80A |
| Speed | Reads, calculates and prints results for 1 plate in about 2 minutes |
| Display | 24 character x 2 line liquid crystal display (LCD) |
| Keyboard | 29 key, enunciating, membrane switch |
| Calculation Modes | Absorbance only, single point calibration from astandard, cutoff, % absorbance multi-point curve fit, point to point curve fit, best fit linear, ln, loglogit, and quadratic |
| Other features | Mixer, real-time clock, self check system, controls locator, partial plate reading, complete user prompting, flags and error messages, cutoff comparisons, curve editing and graphing,monochromatic option, read format 8 vs 12 direction |
| Internal Printer | Thermal, dot matrix, 20 characters per line |
| Serial Port | Thermal, dot matrix, 20 characters per line |
| Power Source | Switch selectable power supply (115V or 230V indicated) Voltage Source 110-120/220-240 VAC from 50 to 60 Hz, CAT II
Power consumption is less than 50 Watts Fuse: two 1/2A, T rating, 250V, 3AG type fuse |
| Certifications and Compliances | Listed to UL Standard 1262, Laboratory Equipment CE Certified of Conformity: Conforms to EN 50082-1 and EN 55022 standards, following the provisions of the 89/336/EEC directive |
| PHYSICAL | |
| Enclosure | Fire-retardant ABS Plastic cover with metal base |
| Dimensions | Approx. 43 x 37 x 18 cm |
| Weight | 11.4 kg |