The Awareness Technology 2600 Microplate Re performs fully automatic reading- The 2600 Microstrip Reader uses biochromatic optics. The standard model includes 4 filters with the option of 6 filters at a nominal charge.
On board data reduction calculations include regressions- curve fitting and cutoff absorbance calculations with step-by-step user prompting and the large non volatile memory can store up to 36 user tests.
Communication with an external device can be completed using the included serial and parallel ports.
| System | |
| Display | Alpha numeric- 16 character LCD |
| Keyboard | 16-key- domed membrane switch- enunciating |
| Power Requirements | 115V or 230V AC- 50-60Hz (switch selectable) |
| Serial Port | Output only- custom cable available- 2400 baud- 1 start- 8 data- 1 stop- no parity- no handshake |
| Processor | Z80A or Z180 |
| Non-volatile Memory | Battery supported RAM- holds more than 32 tests and curves |
| Photometer | |
| Linear Measurements | -0.20 to 3.0 Absorbance units (A) |
| Photometric Accuracy | +/- 1 percent or better |
| Stability | Drift of no more than 0.005A in 8 hours |
| Light Source | Tungsten lamp with lamp saver feature |
| Standard Wave Lengths | 340- 405- 450- 505- 545- 600nm |
| Filter Types | Multiple cavity hard coat interference. 10nm half bandpass |
| Vessel | Standard 96-well microtiter plates- or strip trays |
| Dimensions | |
| L x W x D inch | 9 x 13.25 x 4.875 |
| Weight | 10 Ilbs |